DQRID : D080313.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/02/2008223003/10/20082130Incorrect
more
Subject:
TWP/SKYRAD/C2 - Direct Normal reading too low
DataStreams:twpskyrad20sC2.a0, twpskyradC2.00, twpskyrad60sC2.b1
Description:
The NIP (direct_normal shortwave irradiance) measurement was reading too low during this 
time period.  No TWP Ops documentation could be found to determine the cause but data 
review indicate that the NIP appeared to be reading low by about 25W/m2 during this period.  
The NIP data had well defined steps in irradiance when compared to three component checks 
indicating the problem period.
Suggestions: 
Data is not recoverable but can be re-created using the component sum method where 

for each 1 minute record:
Direct = (Global - Diffuse)/cos(zenith angle),
where Direct = short_direct_normal
Global = down_short_hemisp
Diffuse = down_short_diffuse_hemisp

See SIRS instrument handbook for details.
Measurements:twpskyrad60sC2.b1:
  • short_direct_normal_min
  • short_direct_normal
  • short_direct_normal_std
  • short_direct_normal_max