DQRID : D090225.16
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/10/2008000009/19/20082359Incorrect
more
Subject:
HFE/AERI/M1 - Reprocessed: Recalibration
DataStreams:hfeaeriM1.00
Description:
Note to Reprocessing Team:

Per P090127.1, two of the three thermistors that are used to monitor the
temperature of the ambient blackbody started reporting anomalous values (i.e., way outside 
normal range).  The calibration software uses a weighted average of these thermistors to 
determine the radiating temperature of the blackbody; therefore the calibration 
temperatures were off.  The problem was identified remotely and the weighting was changed on 19 
September to use only the one good temperature measurement. 

The HFE.M1 AERI netcdf data for 20080910-20080919 have been reprocessed to apply corrected 
calibrations.  These data were approved for archival 20090225.

The raw HFE.M1 AERI data (hfeaeriM1.00) were not corrected.  If these data are reprocessed 
from the raw for any reason in the future, the recalibration will need to be re-applied.
Suggestions: 
Measurements:hfeaeriM1.00:
  • Raw data stream - documentation not supported
more