DQRID : D070801.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/19/2007160010/16/20090300Suspect
more
Subject:
SGP/MWR/B1 - 23.8-GHz brightness temperature slightly high
DataStreams:sgpmwrlosB1.b1, sgpmwrtipB1.a1, sgp5mwravgB1.c1
Description:
The radiometer was refurbished and tested at the central facility in 2007. During testing 
the 23.8-GHz brightness temperature was found to be about 1.5 K higher than the 
measurements of the C1 radiometer. 

On 6/19/07 the radiometer was installed at BF1. A comparison with model calculation showed 
that the measured 23.8-GHz brightness temperature is slightly higher than the model. 
This offset will affect the PWV retrieval that is on an average about 1.3 mm higher than the 
radiosonde measurements. The rms error of the retrieved PWV is about 1 mm higher than 
specified in the netcdf file.

On 10/16/2009 the radiometer was removed because of shut down of the boundary facility.