Subject: | SGP/MWR/E14 - Bias in 23.8-GHz measurements |
DataStreams: | sgpmwrtipE14.a1, sgpmwrlosE14.b1
|
Description: | Since the instrument was returned after a power supply failure the 23.8-GHz channel is
constantly 3 K higher than the other co-located radiometer (C1). Since the instrument was
just returned after extensive refurbishing at the manufacturer the cause of the bias is not
clear. |
Suggestions: | The bias in the measurements causes a positive bias in the PWV of ~ 2.5 mm. This bias
should be kept into consideration when using these data. |
Measurements: | sgpmwrtipE14.a1: more
sgpmwrlosE14.b1: more
|