DQRID : D081024.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/20/1999170011/10/20091300Suspect
more
Subject:
SGP/SWATS/E8 - Sensor failure, 35cm W
DataStreams:sgpswatsE8.b1
Description:
The sensor installed at 35 cm at the west profile has produced questionable data ince it 
was installed back in 1996.  Once the sensor began to stabilize at 17:00 GMT on 07/20/99, 
it has produced questionable or incorrect values of temperature rise.  From 07/20/99 to 
the present date, 35cm W temperature rise values have stayed within a range of 
approximately 0.8-2.0 C, while the sensor from the 35cm E profile has produced an approximate range 
of 1.5-4.0 C.

In addition, the calibration for this sensor is so bad it is not possible to correct it 
using field data.  The range of temperature rise values produced by this sensor is too 
limited to reliably re-calibrate the instrument.  Under the old ARM calibration for SWATS, 
the calibrated reference temperature change should fall within a range of 1.45-4.00 C (the 
new range is 1.38-3.96 C).  For this sensor, the reference temperature change has fallen 
within an approximate range of 0.2-1.7 C and the calibration is actually adjusting the 
trise measurements in the wrong direction.  

Therefore, all temperature rise, matric potential and water content data 
are all highly suspect from the 35cm W sensor.

The soil temperature data appear to be OK.

Instrument removed from service 20091110.
Suggestions: 
Use 35cm E measurements
Measurements:sgpswatsE8.b1:
  • trise_W
  • watcont_W
  • soilwatpot_W
more