Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/24/2005 | 0000 | 10/14/2009 | 1100 | Missing |
Subject: | SGP/SWATS/E1 - Sensor failure, 85cm E |
DataStreams: | sgpswatsE1.b1 |
Description: | Both trise and tsoil for the 85cm E sensor dropped out reporting values of -9999 starting on 08/24/2005. Interestingly, both this sensor and the one at 85 cm W (D040730.1) began to report data again June-August 2006. Both sensors have reported values of -9999 after this brief period through the current date. This problem impacts tsoil_E, trise_E, soilwatpot_E, and watcont_E. Instrument was removed from service 20091014. |
Suggestions: | No 85cm level data available after 20050824. |
Measurements: | sgpswatsE1.b1: more
|