Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/09/2008 | 0100 | 01/09/2009 | 0300 | Does not affect quality |
Subject: | TWP/MWR/C2 - Incorrect ambient temperature during rain |
DataStreams: | twpmwrtipC2.a1, twpmwrlosC2.b1 |
Description: | The ambient temperature was approximately 15 degree higher than what it should be when it was raining and the heater came. On Jan 9 2009 the dewblower was replaced which solved the problem. The increased ambient temperature (tkair) was happening during times of rain and users should always disregard MWR data during rain events. Therefore this DQR is colored transparent. |
Suggestions: | |
Measurements: | twpmwrtipC2.a1: more
twpmwrlosC2.b1: more |