DQRID : D090407.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/13/2009000004/15/20091950Does not affect quality
more
Subject:
NSA/MWRP/C1 - Reprocessed: Negative bias in LWP
DataStreams:nsamwrpC1.b1
Description:
After the radiometer was installed at the site following extensive refurbishing the 
retrieved LWP had a large negative bias (~-0.07 mm). The calibration coefficients were modified 
and that resolved the problem. Data for February, March and part of April were 
reprocessed with the revised calibrations and archived in July.
Suggestions: 
Measurements:nsamwrpC1.b1:
  • brightnessTemperature
  • totalPrecipitableWater
  • totalPrecipitableWater2
  • liquidWaterPath
  • liquidWaterPath2
more