DQRID : D090507.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/22/2009092505/16/20091010Suspect
more
Subject:
GRW/CEIL/M1 - Receiver failure
DataStreams:grwceilM1.b1
Description:
The AMF VCEIL had lower-than-normal laser transmitter power, ~84%, so the instrument was 
replaced with a spare on 4/22. The spare VCEIL had just been repaired but began to display 
receiver failure warnings.

Diagnostic information was sent to Vaisala and we discussed either replacing the receiver 
with the one from the original VCEIL (with the degraded transmitter) or replacing the 
transmitter in the original VCEIL with the one from the spare. Vaisala recommended the 
latter because the original VCEIL is a much newer unit.

On 5/16, the site operator replaced the transmitter in the original VCEIL with the one 
from the spare, realigned the optics, recalibrated the laser, and put the instrument back 
into service, successfully correcting the problem.

The data during this period were intermittently incorrect, containing warning flags and 
alarms and unreasonable cloud base heights.
Suggestions: 
Do not use the data when status_flag = "A" indicating an alarm.
Measurements:grwceilM1.b1:
  • backscatter
  • vertical_visibility
  • second_cbh
  • third_cbh
  • alt_highest_signal
  • first_cbh
  • background_light
more