Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/05/2009 | 0000 | 09/01/2009 | 0000 | Suspect |
Subject: | SGP/MWR/B1 - Increased noise level |
DataStreams: | sgpmwrlosB1.b1, sgpmwrtipB1.a1 |
Description: | The radiometer was experiencing intermittent increase in the noise level of the two channels. Although it looks like the radiometer was still able to calibrate, the increased noise level may have resulted in a reduced number of tip curves collected. The increased noise was intermittent and lasted for a few hours at a time. |
Suggestions: | Disregard the short periods of time when the high noise appears. |
Measurements: | sgpmwrtipB1.a1: more
sgpmwrlosB1.b1: more |