DQRID : D090515.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/09/2009000008/31/20091819Incorrect
more
Subject:
SGP/SMOS/E1 - RH high when near saturated
DataStreams:sgp30smosE1.b1, sgp1440smosE1.b1, sgp1smosE1.b1
Description:
Instrument RH values consistently exceeded MAX and did not compare well to surrounding 
sites.  Problem was apparent only when local RH values were in excess of 90%, usually 
between 02-16Z.
Suggestions: 
Use nearby MesoNet or SMOS RH data.
Measurements:sgp30smosE1.b1:
  • sd_rh
  • qc_vap_pres
  • qc_rh
  • rh
  • qc_sd_rh
  • qc_sd_vap_pres
  • vap_pres
  • sd_vap_pres
more
sgp1440smosE1.b1:
  • max_vap_pres
  • time_max_rh
  • min_vap_pres
  • max_rh
  • time_min_rh
  • min_rh
  • time_min_vap_pres
  • time_max_vap_pres
more
sgp1smosE1.b1:
  • rh
  • vap_pres
  • qc_vap_pres
  • qc_rh
more

Close this window