Description: | The SKYRAD IRT was used in testing a revised serial communication program. During the
testing, the analog output configuration was to changed to default values (used at TWP and
NSA sites) that are different than those used at AMF. So, the IRT's analog output was set
to a scale of 25K/V and an offset of 213.2K but the logger converted the signal using a
scale of 26K/V and an offset of 173.2K.
The sky temperature measurements were corrected using the following equation
Tnew = (25/26)(Told-173.2K)+213.2K
and the data were reprocessed. The reprocessed data were archived in April 2011. |