DQRID : D090923.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/10/2009135710/28/20091315Incorrect
more
Subject:
SGP/SIRS/E3 - Multiplexer problem
DataStreams:sgpsirsE3.b1, sgpsirs20sE3.a0
Description:
The down_short_hemisp, down_long_shaded_hemisp and up_long_hemisp values drop out and 
spike during this time period due to a failing multiplexer. The problem was progressive and 
after 9/19 the mux had completely failed and those values go completely missing.  

The problem was not resolved since the site was scheduled for decommissioning. Preventive 
maintenance was ended as of Oct 1 2009. The instrument was removed from service 20091028.
Suggestions: 
Data are not recoverable and are bad until the end of the datastream.
Measurements:sgpsirs20sE3.a0:
  • inst_up_long_dome_resist
  • inst_up_long_hemisp_tp
  • inst_global
  • inst_down_long_shaded_dome_resist
  • inst_down_long_shaded_case_resist
  • inst_up_long_case_resist
  • inst_down_long_hemisp_shaded_tp
more
sgpsirsE3.b1:
  • up_long_hemisp_max
  • inst_up_long_dome_temp
  • down_long_hemisp_shaded_min
  • inst_up_long_case_temp
  • up_long_hemisp_std
  • down_short_hemisp_max
  • down_short_hemisp_std
  • down_long_hemisp_shaded_max
  • down_long_hemisp_shaded_std
  • up_long_hemisp_min
  • up_long_netir
  • down_short_hemisp
  • down_short_hemisp_min
  • inst_down_long_shaded_case_temp
  • inst_down_long_shaded_dome_temp
  • down_long_hemisp_shaded
  • down_long_netir
  • up_long_hemisp
more

Close this window