DQRID : D091112.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/05/2009180911/25/20092000Incorrect
more
Subject:
SGP/MFR10M/C1 - Bad connection
DataStreams:sgpmfr10mC1.b1, sgpmfr10mC1.a0
Description:
The instrument was pulled for quarterly calibration on 20091105.  When the instrument was 
reinstalled, the head cable was not making a good connection to the head.  Cleaning the 
connector and reinstalling resolved the issue.