Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/05/2009 | 1800 | 12/02/2009 | 2115 | Incorrect |
Subject: | SGP/MFRSR/E11 - Alignment and heater board |
DataStreams: | sgpmfrsrlangplotE11.c1, sgpmfrsrE11.b1, sgpmfrsrlangleyE11.c1, sgpmfrsraod1michE11.c1, sgpmfrsrE11.a0 |
Description: | Instrument had a shading issue since reinstallation after calibration. Additionally, the heater controller failed on 20091121. Prior to the heater controller failure, data from sunrise until 1800 UTC were fine. After the failure, no data should be used. Maintenance on 20091202 resolved both the alignment and heater controller issues. |
Suggestions: | |
Measurements: | sgpmfrsrlangplotE11.c1:
sgpmfrsrE11.b1:
sgpmfrsrE11.a0:
sgpmfrsrlangleyE11.c1:
sgpmfrsraod1michE11.c1:
|