Subject:
SGP/QCRAD/E27 - Reprocess: Incorrect calibration coefficients
DataStreams:sgpqcrad1longE27.c2, sgpqcrad1longE27.s2, sgpqcrad1longE27.c1, sgpqcrad1longE27.s1
Description:
Logger reverted to previous years configuration, possibly due to power outage and 
non-current logger program in storage module.  Note, the tracker had failed during this period as 
well.

The configurations are as follows:
Correct S/N and coefficients for this station during this time
Diffuse PSP:   33268F3
PSP-US:        30811F3
NIP:	       29578E6
PSP-DS:        30903F3
Diffuse PSP: 118.49 W/(m^2*mV)
PSP-US:      116.66 W/(m^2*mV)
NIP:	     123.30 W/(m^2*mV)
PSP-DS:      128.92 W/(m^2*mV)

Incorrect S/N and coefficients applied during this time range.
Diffuse PSP:   33366F3
PSP-US:        30961F3
NIP:	       29009E6
PSP-DS:        20386F3
Diffuse PSP: 111.68 W/(m^2*mV)
PSP-US:      111.85 W/(m^2*mV)
NIP:	     111.03 W/(m^2*mV)
PSP-DS:      104.15 W/(m^2*mV)

Data must be reprocessed for all shortwave instruments; longwave instruments are 
unaffected. The following corrections must be applied:

down_short_diffuse_hemisp(new) = down_short_diffuse_hemisp(old) * 118.49/111.68
up_short_hemisp(new) = up_short_hemisp(old) * 116.66/111.85
short_direct_normal(new) = short_direct_normal(old) * 123.30 / 111.03
down_short_hemisp(new) = down_short_hemisp(old) * 128.92 / 104.15

Reference instrument DQR D090521.1
Suggestions: 
Recalculate shortwave flux data as described in the discussion.
Measurements:sgpqcrad1longE27.s2:
  • BestEstimate_down_short_hemisp
  • up_short_hemisp
  • short_direct_normal
  • down_short_diffuse_hemisp
more
sgpqcrad1longE27.c2:
  • short_direct_normal
  • BestEstimate_down_short_hemisp
  • down_short_diffuse_hemisp
  • down_short_hemisp
  • up_short_hemisp
more
sgpqcrad1longE27.s1:
  • down_short_diffuse_hemisp
  • up_short_hemisp
  • BestEstimate_down_short_hemisp
  • short_direct_normal
more
sgpqcrad1longE27.c1:
  • down_short_diffuse_hemisp
  • BestEstimate_down_short_hemisp
  • up_short_hemisp
  • down_short_hemisp
  • short_direct_normal
more

Close this window