DQRID : D131231.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/18/2005000001/21/20092220Suspect
more
Subject:
SGP/IRT/E5 - Sky temperature bias
DataStreams:sgpirtE5.b1, sgpirt200msE5.a1
Description:
Measurements of sky temperature, especially those <250K, were positively biased compared 
with AERI because there was no compensation for the reflectivity of the gold mirror. The 
problem was corrected by changing the IRT emissivity from 1.000 to 0.987.

Use sky temperature measurements less than 230K with caution. Biases as large as 30K have 
been observed when the actual temperature, as measured by AERI, is less than 210 K.
Suggestions: 
Measurements:sgpirt200msE5.a1:
  • sky_ir_temp
more
sgpirtE5.b1:
  • sky_ir_temp
  • sky_ir_temp_max
  • sky_ir_temp_min
more

Close this window