Subject: | SGP/CCN/C1 - CCN signal is low from possible leak and decline in laser power |
DataStreams: | sgpaosccnavgC1.c1, sgpaosccn100C1.a1, sgpaosccnavgC1.c2, sgpaosccnC1.a1, sgpaosccnC1.00
|
Description: | CCN was returned with new laser. Signal was unusually low and flow calibration had changed
substantially. A leak is suspect as well as declining laser power as new laser was
defective. |
Suggestions: | Signal is low and not trusted. Use with caution. |
Measurements: | sgpaosccnC1.a1: - CCN_FirstStageMonitorVoltage
- CCN_ProportionalValveVoltage
- CCN_ss_set
- CCN_dT_TEC3_TEC1_StdDev
- CCN_T_sample
- CCN_P_sample
- CCN_Q_sample
- N_CCN_dN
- CCN_temp_unstable_flag
- CCN_Temperature_Gradient
- N_CCN_bin_number
- CCN_T_TEC1
- CCN_T_TEC2
- CCN_T_TEC3
- N_CCN
- CCN_ss_calc
- CCN_T_OPC
- CCN_laser_current
- CCN_Q_sheath
- CCN_T_inlet
- CCN_UpperSizeLimit
- CCN_T_nafion
more
sgpaosccnC1.00: - Raw data stream - documentation not supported
more
sgpaosccn100C1.a1: - CCN_T_nafion
- CCN_temperature_gradient
- CCN_T_sample
- CCN_super_saturation_calculated
- CCN_first_stage_monitor_voltage
- N_CCN_dN
- CCN_P_sample
- N_CCN_bin_number
- CCN_t_read_TEC3
- CCN_t_read_TEC2
- CCN_t_read_TEC1
- CCN_temp_unstable_flag
- CCN_T_inlet
- CCN_ss_set
- N_CCN
- CCN_temperature_std
- CCN_Q_sample
- CCN_proportional_valve_voltage
- CCN_Q_sheath
- CCN_laser_current
- CCN_T_OPC
more
sgpaosccnavgC1.c2: - N_CCN_dN
- N_CCN
- N_CCN_std
- N_CPC
- supersaturation_calculated
- column_temperature_gradient_measured_std
- supersaturation_setpoint
- droplet_size
- column_temperature_gradient_setting
- N_CPC_std
- column_temperature_gradient_measured
more
sgpaosccnavgC1.c1: - N_CPC_std
- column_temperature_gradient_setting
- supersaturation_setpoint
- N_CCN_std
- column_temperature_gradient_measured_std
- N_CCN_dN
- column_temperature_gradient_measured
- N_CCN
- supersaturation_calculated
- N_CPC
- droplet_size
more
|