DQRID : D100716.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/11/2010180008/09/20101400Incorrect
more
Subject:
GRW/MWR/M1 - Incorrect instrument computer set-up
DataStreams:grwmwrlosM1.b1, grwmwrtipM1.a1
Description:
Starting late on 07/11 the instrument began to have very brief (< 1 min) data outages 
every few hours.  By 07/12 data had become noticeably incorrect, particularly the PWV, LWP, 
TB23, and TB31.  At ~17 Z on 0713 the missing data / no data available flags became more 
frequent.

The problem was identified as improper settings of the instrument computer and was solved 
on 8/9.
Suggestions: 
Use MWRP data during specified time
Measurements:grwmwrtipM1.a1:
  • r23
  • liqtip
  • tbsky31tip
  • tc31
  • tnd23I
  • tnd31I
  • tbsky23tip
  • tnd31
  • tnd23
  • tc23
  • tnd_nom23
  • vaptip
  • tnd_nom31
  • r31
more
grwmwrlosM1.b1:
  • tnd_nom31
  • tbsky31
  • liq
  • tbsky23
  • tc31
  • tc23
  • tnd31
  • tnd_nom23
  • tnd23
  • vap
more

Close this window