DQRID : D110418.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/18/2011000009/20/20112359Does not affect quality
more
Subject:
SGP/MWRHF/C1 - Reprocessed: Rapid drift in 90 GHz noise diode
DataStreams:sgpmwrhfC1.b1
Description:
The noise diode of the 90 GHz channel was showing rapid drifts that affected the accuracy 
of the calibration. Although the algorithm is designed to follow slow drifts in the noise 
diode, in this case (because the variation was fast) the computed brightness temperature 
was probably about 4-5 K colder than it should be.

Ingest was modified and data were reprocessed and archived in September 2011
Suggestions: 
Measurements:sgpmwrhfC1.b1:
  • tbsky90
more

Close this window