DQRID : D110421.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/11/2011000001/27/20111920Suspect
more
Subject:
SBS/MWR3C/M1 - Intermittent thermal instability
DataStreams:sbsmwr3cM1.b1
Description:
The radiometers reset the thermal control every time the software was started at 00:00 
UTC. It took about an hour to reestablish the thermal control of the instrument after each 
reset. On Jan 27 a workaround was introduced to avoid this behavior. The software will be 
modified at the next release.
Suggestions: 
Measurements:sbsmwr3cM1.b1:
  • tbsky23
  • tbsky89
  • tbsky30
more