Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/01/2011000010/13/20112200Incorrect
more
Subject:
NSA/GVRP/C1 - High noise in the data
DataStreams:nsagvrpC1.b1
Description:
All channels of the radiometer became suddenly very noisy. The receiver noise figure was 
computed and was found anomalous. Data are unusable during the indicated times. On 10/13 
the instrument was removed and will be sent to the vendor for repairs.
Suggestions: 
Do not use.