Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/01/2011 | 0000 | 10/13/2011 | 2200 | Incorrect |
Subject: | NSA/GVRP/C1 - High noise in the data |
DataStreams: | nsagvrpC1.b1 |
Description: | All channels of the radiometer became suddenly very noisy. The receiver noise figure was computed and was found anomalous. Data are unusable during the indicated times. On 10/13 the instrument was removed and will be sent to the vendor for repairs. |
Suggestions: | Do not use. |
Measurements: | nsagvrpC1.b1: more
|