DQRID : D111118.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/24/2011080010/24/20111400Incorrect
10/23/2011050010/23/20111600Incorrect
10/11/2011030011/11/20111400Incorrect
more
Subject:
SGP/MWRHF/C1 - Dew formation on window
DataStreams:sgpmwrhfC1.b1
Description:
During the indicated time there was moisture or dew on the radiometer window and the 
brightness temperatures are too high. The presence of dew can be recognized by a smooth 
increase in the brightness temperature followed by a smooth decrease over several hours.
Suggestions: 
Users should disregard the affected data
Measurements:sgpmwrhfC1.b1:
  • tbsky150
  • tbsky90
more

Close this window