DQRID : D120420.8
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/12/2011000006/24/20111740Incorrect
more
Subject:
SGP/QCRAD/C1 - NIP failure
DataStreams:sgpqcrad1longC1.c2, sgpqcrad1longC1.s2, sgpqcrad1longC1.s1, sgpqcrad1longC1.c1
Description:
The short_direct_normal (NIP instrument) cable failed during this time period.	Data are 
not recoverable.

Users can derive an estimate of direct normal values using the following formula:
short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle)

Documented in instrument DQR D110628.2
Suggestions: 
Use direct normal from E13, MFRSR, or calculate it from SW global and diffuse as described.
Measurements:sgpqcrad1longC1.c1:
  • short_direct_normal
more
sgpqcrad1longC1.s1:
  • short_direct_normal
more
sgpqcrad1longC1.c2:
  • short_direct_normal
more
sgpqcrad1longC1.s2:
  • short_direct_normal
more

Close this window