DQRID : D120131.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/12/2011000001/25/20122359Suspect
more
Subject:
GAN/CSPHOT/M1 - Instrument problem, Obstruction, collimator
DataStreams:gancsphotaotfiltM1.a1, gancsphotaotM1.a1
Description:
During this time period, there were obstructions in the collimator tubes, causing 
"InGaAs/Silicon Discrepancy" errors. Webs were found. The blockages caused discrepancies in the 
data between the 1640 and the other channels.
Suggestions: 
During this time period mentor recommends using the quality assured level 2.0 AOT 

data from AERONET when available. Level 2.0 data and post-calibration level 1.5 data should 
become available in 2012.  Can also use MFRSR data if available.