Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/01/2012 | 0000 | 01/09/2012 | 0000 | Incorrect |
Subject: | GAN/VDIS/M1 - Diameter_min incorrect |
DataStreams: | ganvdisM1.b1 |
Description: | The diameter_min values failed the valid max of 10.00 mm. Values were flatlined at 100 mm. |
Suggestions: | |
Measurements: | ganvdisM1.b1: more
|