DQRID : D120220.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/04/2012000001/08/20122359Incorrect
01/31/2012000002/01/20122359Incorrect
02/07/2012210002/08/20120500Incorrect
01/12/2012000001/20/20120000Incorrect
more
Subject:
GAN/MWR3C/M1 - Radiometer instability
DataStreams:ganmwr3cM1.b1
Description:
The radiometer showed intermittent instability in the 30 GHz and 89 GHz channel. There was 
a calibration shift between Jan 4 and Jan 8 and a calibration instability between Jan 12 
and Jan 20. It may be possible to recalibrate the data when a recalibration algorithm is 
developed.
Suggestions: 
Users should disregard data during the indicated time as they are incorrect
Measurements:ganmwr3cM1.b1:
  • tbsky89
  • pwv
  • lwp
  • tbsky30
more