Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/09/2012 | 0000 | 02/22/2012 | 0000 | Incorrect |
Subject: | SGP/MWR3C/C1 - Incorrect or missing data |
DataStreams: | sgpmwr3cC1.b1 |
Description: | The 89-GHz receiver started to reset itself continuously. The problem was initially identified with a wrong cable connection, but it was later found that one pin in the power connector of the receiver was broken. The pin was pulled out to temporarily start the radiometer. After the receiver was replaced it was discovered that there was an additional problem with water infiltration in the mini j-box. The mini-jbox was removed and sent for repair. |
Suggestions: | Data during the indicated time are very intermittent and/or corrupted. Users should disregard the data. |
Measurements: | sgpmwr3cC1.b1: more
|