Description: | See below for details of reprocessing that was completed on 20141024. Original
DQR category: 4 (Red, Incorrect).
-----------------------------------------------------------
The calibration coeff and serial number for the down_long_hemisp_shaded was incorrect
during the period specified.
Correct Instrument serial number = 30010F3
K1 = 0.3125
Incorrect Instrument serial number = 30690F3
K1 = 0.2762
Data will be reprocessed.
To correct the one minute data b1 datastream follow the procedure below:
1) Obtain the following variables from the 20s datastream for each one minute of the
affected
time range:
inst_down_long_shaded_dome_resist (Rd)
inst_down_long_shaded_case_resist (Rc)
inst_down_long_hemisp_shaded_tp (tpDIR_mV)
2) using the 20s tp voltage in mV convert to uV.
tpDIR[uV] = inst_down_long_hemisp_shaded_tp[mV]*1000[uV/mV]
3) For each 20s data entry convert Rd and Rc to temperature
Td =1/ (a0 + (a1*LN (Rd*1000) + (a2*LN (Rd*1000)^3)))
Tc = 1/ (a0 + (a1*LN (Rc*1000) + (a2*LN (Rc*1000)^3)))
where a0 = 0.0010295
a1 = 0.0002391
a2 = 0.0000001568
4) Apply for each 20s data entry: (correct k1=0.3125)
WinDIR(W/m^2)= [0.3125*tpDIR] + Sigma*Tc^4 -4*Sigma*(Td^4 - Tc^4)
where,
WinDIR = Corrected down_long_hemisp_shaded (W/m^2)
Sigma = 5.6704E-8 (Stefan-Boltzman Constant)
5) Generate new 1 minute average from the three 20-second measurements |