DQRID : D120521.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/13/2012182204/25/20121608Does not affect quality
more
Subject:
NSA/MWRP/C1 - Reprocessed: Incorrect calibration
DataStreams:nsamwrpC1.b1
Description:
The LN2 calibration was only partially successful, some the K-band channels did not 
calibrate properly. On 4/25 new coefficients were derived from tip curves and were uploaded. 

The data for 4/13-4/25 were reprocessed to apply the 4/25 calibrations.  Reprocessed data 
were archived in July 2012.
Suggestions: 
Measurements:nsamwrpC1.b1:
  • liquidWaterContent
  • waterVaporMixingRatio
  • virtualTemperature
  • brightnessTemperature
  • totalPrecipitableWater
  • totalPrecipitableWater2
  • waterVaporDensity
  • liquidWaterPath
  • liquidWaterPath2
more