Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/01/2008 | 0000 | 07/02/2012 | 2045 | Suspect |
Subject: | SGP/SURTHREF/C1 - Sensors out of calibration |
DataStreams: | sgpsurthrefC1.b1 |
Description: | In Nov. 2011, RH values were noticed to be out of agreement with each other. It was found that the sensors had not been through annual calibrations (some had not been calibrated since 2008). Beginning Dec. 2011, sensors were swapped and calibrated as needed. As of 7/2/2012, all 6 sensors are up to date on calibrations. V2 replaced 12/12/2011. V1/V3/R2 replaced as of 01/24/2011. R1 replaced 4/23/2012. R3 replaced 7/3/2012. |
Suggestions: | No workaround, but keep calibration dates in mind when scrutinizing data disagreements. |
Measurements: | sgpsurthrefC1.b1: more
|