Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/13/2011 | 1243 | 03/27/2012 | 0800 | Incorrect |
Subject: | PGH/AOS/M1 - Humidifier heater stopped working |
DataStreams: | pghaosnephwetM1.a1 |
Description: | The PID controller for the humidograph heater stopped working. The RH inside the wet neph stopped ramping up and down each hour and so there aren't any frh data for the fRH VAP. |
Suggestions: | |
Measurements: | pghaosnephwetM1.a1: more
|