Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/11/2011 | 1800 | 02/10/2012 | 0600 | Suspect |
Subject: | PGH/AOS/M1 - Trailer temp too low, CCN affected |
DataStreams: | pghaosccn100M1.a1 |
Description: | The trailer temperature wasn't regulating very well. Night time temperatures fell to 10 C. The CCN doesn't operate well at low temperatures. As a consequence the instrument temperatures were below the acceptable range for the model calculation of %ss. The instrument still reported a %ss values, but this value is likely off by as much as 0.1%, especially at night when the trailer temperatures were low. |
Suggestions: | Use night time data with caution and understand that they have a higher uncertainty. |
Measurements: | pghaosccn100M1.a1: more
|