DQRID : D120807.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/11/2011180002/10/20120600Suspect
more
Subject:
PGH/AOS/M1 - Trailer temp too low, CCN affected
DataStreams:pghaosccn100M1.a1
Description:
The trailer temperature wasn't regulating very well. Night time temperatures fell to 10 C. 
The CCN doesn't operate well at low temperatures. As a consequence the instrument 
temperatures were below the acceptable range for the model calculation of %ss. The instrument 
still reported a %ss values, but this value is likely off by as much as 0.1%, especially 
at night when the trailer temperatures were low.
Suggestions: 
Use night time data with caution and understand that they have a higher uncertainty.
Measurements:pghaosccn100M1.a1:
  • CCN_ss_set
more

Close this window