Subject: | SGP/SIRS/E38 - Upwelling LW cable failure |
DataStreams: | sgpsirsE38.b1, sgpsirs20sE38.a0
|
Description: | The up_long_hemisp shows large steps and offsets in values (greater than physically
possible incoming radiation) that were eventually attributed to a bad instrument cable
connection.
The up_long_hemisp values during the time range specified were intermittant, varying from
10 to 100% of any given day. |
Suggestions: | Data are not recoverable. |