Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/25/2012 | 0000 | 08/28/2012 | 2045 | Incorrect |
Subject: | PVC/SKYRAD/M1 - Intake Fan Failure |
DataStreams: | pvcskyrad20sM1.a0, pvcskyrad60sM1.b1 |
Description: | PSP (down_short_hemisp) ventilator fan failed from start of PVC deployment. Site ops repaired/replaced on 8/28. Instrument data are not recoverable. The down_short_hemisp data are not recoverable but can be re-constructed using the short_direct_normal and down_short_diffuse_hemispheric measurements and the following formula: Global = Direct * cos(zenith angle) + Diffuse |
Suggestions: | |
Measurements: | pvcskyrad60sM1.b1: more
pvcskyrad20sM1.a0: more |