DQRID : D120918.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/08/2012191008/12/20120940Suspect
more
Subject:
TWP/IRT/C1 - Data acquisition problem
DataStreams:twpirt200msC1.a1, twpirtC1.b1
Description:
Serial communication errors resulted in incomplete raw data resulting in incorrect 
processed data records. Response strings from the IRT were occasionally not parsed correctly 
causing a much smaller value of temperature (for example, '29' instead of '294.6') to be 
recorded.

The acquistion program was updated to improve communication with the IRT.
Suggestions: 
Measurements:twpirtC1.b1:
  • sky_ir_temp_max
  • sky_ir_temp_std
  • sky_ir_temp_min
  • sky_ir_temp
more
twpirt200msC1.a1:
  • sky_ir_temp
more