Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/03/2012 | 1320 | 08/02/2012 | 2350 | Suspect |
Subject: | TWP/IRT/C3 - Data acquisition problem |
DataStreams: | twpirtC3.b1, twpirt200msC3.a1 |
Description: | Serial communication errors resulted in incomplete raw data resulting in incorrect processed data records. Response strings from the IRT were occasionally not parsed correctly causing a much smaller value of temperature (for example, '29' instead of '294.6') to be recorded. The acquistion program was updated to improve communication with the IRT. |
Suggestions: | |
Measurements: | twpirtC3.b1: more
twpirt200msC3.a1: more |