DQRID : D121018.8
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/02/2012153010/18/20121930Incorrect
more
Subject:
SGP/SIRS/E12 - Shadowband misalignment
DataStreams:sgpsirs20sE12.a0, sgpsirsE12.b1
Description:
The E12 tracker had failed from 7/20 to 8/2. The tracker was replaced and returned to 
service on 8/2.  However, the tracker was not aligned correctly until 10/18.  

Data for short_direct_normal (NIP instrument) and down_short_diffuse (8-48 instrument) are 
bad and not recoverable during this entire time range.
Suggestions: 
Measurements:sgpsirs20sE12.a0:
  • inst_direct_normal
  • inst_diffuse
more
sgpsirsE12.b1:
  • short_direct_normal
  • down_short_diffuse_hemisp
  • short_direct_normal_min
  • down_short_diffuse_hemisp_max
  • down_short_diffuse_hemisp_std
  • down_short_diffuse_hemisp_min
  • short_direct_normal_max
  • short_direct_normal_std
more