DQRID : D121024.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/18/2012130010/22/20121259Incorrect
more
Subject:
PVC/SKYRAD/M1 - IRT scale factor
DataStreams:pvcskyrad20sM1.a0, pvcskyrad60sM1.b1
Description:
The IRT was replaced with a unit with the wrong analog output configuration, such that the 
sky temperature was below the measurable range, resulting in a constant output of 0 
volts or 173.2 K.
Suggestions: 
Measurements:pvcskyrad60sM1.b1:
  • sky_ir_temp_max
  • sky_ir_temp_min
  • sky_ir_temp_std
  • sky_ir_temp
more
pvcskyrad20sM1.a0:
  • inst_sky_ir_signal
more