Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/17/2012181709/26/20121839Does not affect quality
more
Subject:
SGP/SIRS/E33 - Reprocessed: Incorrect DIR calibration
DataStreams:sgpsirsE33.b1, sgpsirs20sE33.a0
Description:
During troubleshooting of the UIR (up_long_hemisp) the UIR and DIR (down_long_hemsip) 
pyrgeometers were swapped to help isolate an instrument noise issue.  That swap resulted in 
bad DIR data from 8/14-8/17 (documented in D121010.1)

A replacement DIR instrument installed on 8/17 had incorrect calibration coefficients 
applied.  The serial number was correct.

wrong DIR k1 cal = 0.2967
correct DIR k1 cal =  0.3086

Reprocessing completed 20121217.
Suggestions: 
Use reprocessed data.