Subject: | SGP/SIRS/E13 - Upwelling LW hemisp connection |
DataStreams: | sgpsirsE13.b1, sgpsirs20sE13.a0
|
Description: | On 10/25 an intermittent problem began on the up_long_hemispheric (UIR) instrument dome
thermistor which resulted in incorrect or missing calculated flux. Maintenance tightened
connections on 11/21 which appears to have resolved the problem. |
Suggestions: | Use sgpqcrad1longE13.c1 where data are effectively flagged as bad |
Measurements: | sgpsirsE13.b1: - up_long_hemisp
- up_long_hemisp_min
- up_long_hemisp_std
- up_long_hemisp_max
- up_long_netir
- inst_up_long_case_temp
- inst_up_long_dome_temp
more
sgpsirs20sE13.a0: - inst_up_long_dome_resist
- inst_up_long_hemisp_tp
- inst_up_long_case_resist
more
|