DQRID : D121217.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/10/2012000011/15/20122000Incorrect
more
Subject:
NSA/GVRP/C1 -  Calibration drift
DataStreams:nsagvrpC1.b1
Description:
The radiometer showed a progressive drift in calibration starting approximately on 
10/10/2012 and becoming worst in November. The instrument was calibrated with liquid nitrogen on 
11/15/2012
Suggestions: 
Measurements:nsagvrpC1.b1:
  • brightnessTemperature
more