Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/22/2012 | 1223 | 10/04/2012 | 1359 | Suspect |
Subject: | MAG/PRP/M1 - FRSR head temperature error |
DataStreams: | magprpfrsrM1.a0 |
Description: | During the time specified (Leg2) the MFR head temperature sensor was failing intermittently. At times when the head_temp falls outside the normal range of 38 C to 42 C, the heater likely was not maintaining proper operating temperature, and the radiometric calibration is invalid. The FRSR head was replaced for Leg 3 and beyond. |
Suggestions: | Omit any data records if the head temperature is not in the 38-42C temperature range: head_temp < 38 or head_temp > 42. |
Measurements: | magprpfrsrM1.a0: more
|