DQRID : D130307.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/09/2012000010/25/20122300Suspect
more
Subject:
SGP/CCN/C1 - Reprocess: Decline in CCN at low %ss values
DataStreams:sgpaosccn100C1.a1
Description:
The CCN number concentrations at low %ss values started to decline. DAta at low %ss values 
are suspect. 

Reeprocess to apply the instrument calibration done in November 2012 to the data 
Sep/12-Oct/12. 

Better %ss values can be obtained by applying %ss = (dT-2.63)/16.00, where dT = T3-T1.
Suggestions: 
Wait for reprocessed data
Measurements:sgpaosccn100C1.a1:
  • CCN_super_saturation_calculated
  • CCN_ss_set
more