MAG/OZONE/M1 - Incorrect filter
DataStreams:magaoso3M1.a1
Description:
During a filter change, the filter was incorrectly replaced with the non-porous blue 
filter separator (Operator error).  This reduced the flow and pressure in the unit.  Data were 
invalid until the problem was observed and corrected.
Suggestions: 
Measurements:magaoso3M1.a1:
  • ozone_pressure
  • ozone_bench_temp
  • ozone
more