Description: | See below for details of reprocessing that was completed on 20150228. Original
DQR category: 4 (Red, Incorrect).
-----------------------------------------------------------
The CR23X datalogger was replaced with a CR3000 on 2/21/2012. Incorrect serial numbers and
calibration coefficients were in the CR3000 program for down_short_hemisp PSP,
down_short_diffuse_hemisp 8-48, short_direct_normal NIP, and down_long_hemisp PIR.
The correct S/N and coefficients during the affected time period:
down_short_hemisp: PSP- 30812F3; 127.06
down_short_diffuse_hemisp: 8-48- 33238; 109.51
short_direct_normal: NIP- 29851E6; 119.59
down_long_hemisp: PIR- 30357F3; K0 0.00, K1 0.2577, K2 1.00, K3 -4.00, Kr 0.00
The incorrect S/N and coefficients during the affected time period:
down_short_hemisp: PSP- 9996F3; 100.00
down_short_diffuse_hemisp: 8-48- 99993; 100.12
short_direct_normal: NIP- 99995; 100.00
down_long_hemisp: PIR- 99992; K0 0, K1 0.25, K2 1, K3 -4, Kr 0 |