DQRID : D130329.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/30/2013000003/26/20132122Incorrect
more
Subject:
SGP/SIRS/E13 - UIR Instrument problem
DataStreams:sgpsirsE13.b1, sgpsirs20sE13.a0
Description:
Due to a bad connection the up_long_hemisp datastream (upwelling Epply PIR instrument) has 
intermittent missing data during the time range specified.

Missing data are not recoverable.  However, the sgpqcrad1longE13.c1 datastream adequately 
identifies missing values and removes them.  Recommendation to the user is to use this 
VAP datastream for valid values of up_long_hemisp irradiance from this station.
Suggestions: 
Use sgpqcrad1longE13.c1
Measurements:sgpsirsE13.b1:
  • up_long_hemisp
  • up_long_hemisp_min
  • up_long_hemisp_std
  • up_long_hemisp_max
  • up_long_netir
  • inst_up_long_case_temp
  • inst_up_long_dome_temp
more
sgpsirs20sE13.a0:
  • inst_up_long_dome_resist
  • inst_up_long_hemisp_tp
  • inst_up_long_case_resist
more

Close this window