DQRID : D130930.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/01/2013000009/24/20131700Suspect
more
Subject:
SGP/SIRS/E34 - Ventilation fan failure
DataStreams:sgpsirsE34.b1, sgpsirs20sE34.a0
Description:
On June 1, 2013 the ventilator fan on the PSP (down_short_hemisp) instrument failed.  

This adversely affected the down_short_hemisp data during the specified period.  Data 
review indicates the daytime DS data appear unaffected during the problem period. However, 
the additional uncertainly in down_short_hemisp due to this failure is not quantifiable.

The down_short_hemisp from the b1 datastream should be avoided during this time period.  
The sgpqcrad1longE34 BestEstimate_down_short_hemisp remains the most accurate estimate of 
down_short_hemisp since it primarily uses the unaffected direct normal and diffuse 
measurements.
Suggestions: 
Use QCRAD data
Measurements:sgpsirsE34.b1:
  • down_short_hemisp_max
  • down_short_hemisp_std
  • down_short_hemisp_min
  • down_short_hemisp
more
sgpsirs20sE34.a0:
  • inst_global
more