Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/06/2013 | 1800 | 11/16/2013 | 0000 | Missing |
Subject: | SGP/AOS/C1 - Dry nephelometer failure |
DataStreams: | sgpnoaaaosC1.b1, sgpaosnephdryC1.b1, sgpnoaaaosavgC1.b1 |
Description: | There was a communications problem with the nephelometer that resulted in intermittent data transmission and bad background values. The neph was removed for repair on Nov 15th. Use the wet neph data instead during this time as the humidifier was off. After this time the wet neph was put in the dry neph position until the neph could be repaired. |
Suggestions: | Use the wet neph data |
Measurements: | sgpnoaaaosavgC1.b1: more
sgpaosnephdryC1.b1: more sgpnoaaaosC1.b1: more |