Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/21/2013 | 1720 | 12/17/2013 | 1740 | Does not affect quality |
Subject: | SGP/SIRS/E31 - Reprocessed: Incorrect LW calibration coefficients |
DataStreams: | sgpsirsE31.b1 |
Description: | During the time range specified the UIR and DIR serial numbers were correct but incorrect calibration coefficients were applied in the datalogger. up_long_hemisp (SN# 31236F3) wrong k1 cal :0.2725 W/(m^2*uV) correct k1 cal: 0.2740 W/(m^2*uV) down_long_hemisp (SN# 30837F3) wrong k1 cal :0.25407 W/(m^2*uV) correct k1 cal: 0.2584 W/(m^2*uV) |
Suggestions: | Data were reprocessed in February 2014. |
Measurements: | sgpsirsE31.b1: more
|